All six inch (220mm) and 8 inch (300mm) chucks
manufactured by MC Systems, Inc. are typically spring mounted to a an
electrically insulated base by three screws. Adjustment of these three
screws will cause the chuck surface to tilt. Using these three screws, one
can flatten (planarize) the chuck surface to + & - 0.0005”.
First, you must provide an indicator to sense the
position of the surface of the chuck. This may be a dial indicator 1)
mounted firmly to the prober platen or if the system has a high power
microscope, one can use the depth of field 2) at high magnification which
is simple and easy.
Move the chuck until one of the three adjustment screws
is near the center of the platen opening. Using a dial indicator, place
the indicator on the platen with the sensing arm lightly touching the
surface of the chuck. Move the platen up or down until the dial indicator
reads zero. Using the stage adjustments, move the chuck top so that the
next counter clockwise screw adjustment is near the center of the chuck.
If the dial indicator still reads zero then great, if not, use an Allen
wrench in the adjustment screw to bring the dial indicator to zero. Use
the same procedure to move to the third screw, and adjust it, if need be.
There will be inter action between these adjustments, therefore, return to
the starting position of the first setting, if this is not zero, then use
the platen adjust to bring the dial indicator to zero. Repeat the previous
procedure on the next two adjustment screws. The overall procedure may
have to be repeated several times.
2) Move the chuck until one of the three adjustment
screws is near the center of the field of view of the microscope.
depending on the selection of chuck height sensing. Be sure the microscope
is on its highest power as the depth of field will be quite narrow. Focus
the microscope on the chuck top. Using the stage adjustments, move the
chuck top so that the next counter clockwise screw adjustment is near the
center of the chuck. If the microscope is still focused then great, if
not, use an Allen wrench in the adjustment screw to bring the microscope
back into focus. Use the same procedure to move to the third screw, and
adjust it, if need be. There will be inter action between these
adjustments, therefore, return to the starting position of the first
setting, if this is not zero, then refocus the microscope. Repeat the
previous procedure on the next two adjustment screws. The overall
procedure may have to be repeated several times